期刊
ANALYTICA CHIMICA ACTA
卷 680, 期 1-2, 页码 41-47出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2010.09.023
关键词
Thiol; Yeast; Speciation; Orbitrap; Derivatization
资金
- NRC, Canada
- Tubitak
An analytical method was developed for the determination of thiols in biological samples. Reverse phase chromatography coupled to ICP quaduropole MS or Orbitrap MS was employed for the separation and detection of thiols. For the determination of total thiols, oxidized thiols were reduced using dithiothreitol (DTT). Reduction efficiencies for species of interest were found to be close to 100%. Reduced thiols were derivatized by p-hydroxymercuribenzoate (PHMB) and then separated on a C8 column. Optimization of the extraction, separation and detection steps of the HPLC-ICP-MS and HPLC-Orbitrap MS methods was carried out. Detection limits for cysteine, homocysteine, selenocysteine, glutathione, selenomethionine and cysteinyl-glycine were found to be 18,34,39,12,128 and 103 fmol, respectively, using HPLC-Orbitrap MS and 730, 1110, 440, 1110 and 580 fmol for cysteine, homocysteine, selenocysteine, glutathione, and cysteinyl-glycine using HPLC-ICP-MS. Contrary to expectation, the LODs and RSDs are higher for the HPLC-ICP-MS instrument, therefore HPLC-Orbitrap MS was used for the determination of thiols in yeast samples. Three different brands of baker's yeast and a selenized yeast were analyzed. The GSH and cysteine levels found in these samples ranged from 4.45 to 17.87 mu mol g(-1) and 0.61 to 1.32 mu mol g(-1), respectively. (C) 2010 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据