4.7 Article

Exploiting dynamic reaction cell inductively coupled plasma mass spectrometry (DRC-ICP-MS) for sequential determination of trace elements in blood using a dilute-and-shoot procedure

期刊

ANALYTICA CHIMICA ACTA
卷 639, 期 1-2, 页码 13-18

出版社

ELSEVIER
DOI: 10.1016/j.aca.2009.03.016

关键词

Trace elements; Interferences; Inductively coupled plasma mass spectrometry; Direct analysis; Clinical specimens; Dynamic reaction cell

资金

  1. Fundao de Amparo a Pesquisa do Estado de Sao Paulo (FAPESP)
  2. Conselho Nacional de Desenvolvimento Cientifico e Tecnologico (CNPq)

向作者/读者索取更多资源

Inductively coupled plasma mass spectrometry with quadrupole (q-ICP-MS) and dynamic reaction cell (DRC-ICP-MS) were evaluated for sequential determination of As, Cd, Co, Cr, Cu, Mn, Pb, Se, Tl, V and Zn in blood. The method requires as little as 100 mu L of blood. Prior to analysis, samples (100 mu L) were diluted 1:50 in a solution containing 0.01% (v/v) Triton (R) X-100 and 0.5% (v/v) nitric acid. The use of the DRC was only mandatory for Cr, Cu, V and Zn. For the other elements the equipment may be operated in a standard mode (q-ICP-MS). Ammonia was used as reaction gas. Selection of best flow rate of ammonium gas and optimization of the quadrupole dynamic band-pass tuning parameter (RPq) were carried out, using a ovine base blood for Cr and V and a synthetic matrix solution (SMS) for Zn and Cu diluted 1:50 and spiked to contain 1 mu g L-1 of each element. Method detection limits (3 s) for As-75, Cd-114, Co-59, Cr-51, Cu-63 Mn-55, Pb-208, Se-82, Tl-205, V-51, and Zn-64 were 14.0, 3.0, 11.0, 7.0, 280, 9.0, 3.0, 264, 0.7, 6.0 and 800 ng L-1, respectively. Method validation was accomplished by the analysis of blood Reference Materials produced by the National cle Sant Publique du Quebec (Canada). (C) 2009 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据