期刊
ANALYTICA CHIMICA ACTA
卷 623, 期 1, 页码 76-81出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2008.06.010
关键词
bulk-modified screen-printed electrode; bismuth oxide; lead (II); cadmium (II); constant current stripping; chronopotentiometry; soil extracts; wastewater
A bulk-modified screen-printed carbon electrode characterised for metal ion detection is presented. Bismuth oxide (Bi2O3) was mixed with graphite-carbon ink to obtain the modified electrode. The best composition was 2% Bi2O3 (wt%) in the graphite-carbon ink. The modified electrode with onboard screen-printed carbon counter and silver-silver chloride pseudo-reference electrodes exhibited good performance in the electrochemical measurement of lead (II) and cadmium (II). The electrode displayed excellent linear behaviour in the concentration range examined (20-300 mu g L-1) with limits of detection of 8 and 16 mu g L-1 for both lead (II) and cadmium (II), respectively. The analytical utility of the modified electrode was illustrated by the stripping chronopotentiometric determinations of lead (II) in soil extracts and wastewater samples. (C) 2008 Elsevier B.V All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。