期刊
ANALYTICA CHIMICA ACTA
卷 609, 期 2, 页码 160-168出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2008.01.001
关键词
sulfide; inductively coupled plasma-atomic; emission spectroscopy; vapor generator; water analysis
Two new, simple and accurate methods for the determination of sulfide (S2-) at low levels (mu gL(1-)) in aqueous samples were developed. The generation of hydrogen sulfide (H2S) took place in a coil where sulfide reacted with hydrochloric acid. The resulting H2S was then introduced as a vapor into an inductively coupled plasma-atomic emission spectrometer (ICP-AES) and sulfur emission intensity was measured at 180.669 nm. In comparison to when aqueous sulfide was introduced, the introduction of sulfur as H2S enhanced the sulfur signal emission. By setting a gas separator at the end of the reaction coil, reduced sulfur species in the form of H2S were removed from the water matrix, thus, interferences could be avoided. Alternatively, the gas separator was replaced by a nebulizer/spray chamber combination to introduce the sample matrix and reagents into the plasma. This methodology allowed the determination of both sulfide and sulfate in aqueous samples. For both methods the linear response was found to range from 5 mu gL(-1) to 25 mg L-1 of sulfide. Detection limits of 5 jig L-1 and 6 jig L-1 were obtained with and without the gas separator, respectively. These new methods were evaluated by comparison to the standard potentiometric method and were successfully applied to the analysis of reduced sulfur species in environmental waters. (c) 2008 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐