4.7 Article

Optimisation and critical evaluation of a collision cell technology ICP-MS system for the determination of arsenic in foodstuffs of animal origin

期刊

ANALYTICA CHIMICA ACTA
卷 611, 期 2, 页码 134-142

出版社

ELSEVIER
DOI: 10.1016/j.aca.2008.02.018

关键词

inductively coupled plasma mass; spectrometry; collision cell; foodstuffs; arsenic; interferences

向作者/读者索取更多资源

The determination of arsenic (As-75) was studied using an ICP-MS equipped with collision cell technology (CCT). Different mixtures of gases (He and H-2) were tested using HCl conditions and a He flow rate of 4 mL min(-1) was found to be suitable for the removal of the poly-atomic spectral interference [(ArCl)-Ar-40-Cl-35](+). Trueness of the optimised method has been evaluated in both standard and CCT modes on six certified reference materials in foodstuffs of animal origin and on three external proficiency testing schemes (FAPAS). The results obtained generally coincided with the certified values, except for CCT mode in some categories of samples (meat, mussels and milk powder), for which a positive bias on results was observed due to the formation of poly-atomic interferences within the collision cell. The main interferences were studied and their contributions estimated. [(FeOH)-Fe-58-O-16-H-1](+) and [(GeH)-Ge-74-H-1](+) were the most significant interferences formed in the cell. Finally, different parameters (e.g. hexapole and quadrupole bias voltage, nebuliser gas flow) were optimised to try to attenuate these interferences. (C) 2008 Published by Elsevier B.V.