Flow injection determination of aluminium by spectrofluorimetric detection after complexation with N-o-vanillidine-2-amino-p-cresol:: The application to natural waters

期刊

ANALYTICA CHIMICA ACTA
卷 611, 期 1, 页码 62-67

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2008.01.044

关键词

flow injection; aluminium determination; N-o-vanillidine-2-amino-p-cresol; spectrofluorimetry; seawater; river water

向作者/读者索取更多资源

An on-line flow injection spectrofluorimetric method for the direct determination of aluminium in water samples is described. The method is based on the reaction of aluminium with N-o-vanillidine-2-amino-p-cresol (OVAC) in acidic medium at pH 4.0 to form a watersoluble complex. The excitation and emission wavelengths were 423.0 and 553.0 nm, respectively, at which the OVAC-Al complex gave the maximum fluorescence intensity at pH 4.0 in a 50% methanol-50% water medium at 50 degrees C. An interference from fluoride ions was minimised by the addition of Be2+. Other ions were found not to interfere at the concentrations likely to be found in natural waters. The proposed methods were validated in terms of linearity, repeatability detection limit, accuracy and selectivity. Under these conditions, the calibration was linear up to 1000 mu g L-1 (r=0.999). The limit of detection (3 sigma) for the determination of Al(III) was 0.057 mu g L-1 and the precision for multiple determinations of 3 ng mL(-1) Al(III) prepared in ultra-pure water was found to be 0.62% (n=10). The Schiff base ligand could be used to determine ultra-trace aluminium from natural waters. Analysis of environmental certified reference materials showed good agreement with the certified values. The procedure was found to be equally applicable to both fresh-water and saline solutions, including seawater. (c) 2008 Published by Elsevier B.V.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据