期刊
ANNALS OF CLINICAL BIOCHEMISTRY
卷 44, 期 -, 页码 463-466出版社
SAGE PUBLICATIONS INC
DOI: 10.1258/000456307781646102
关键词
-
Background The measurement of the serum concentration of fibroblast growth factor-23 (FGF-23) is beginning to be used as a diagnostic tool in renal phosphate wasting disorders. Having observed an increased serum FGF-23 in three subjects with low circulating ferritin concentrations we investigated the association between low ferritin and raised serum FGF-23. Methods We measured FGF-23 in 150 random anonymized serum samples with ferritin concentrations between < 5 and 50 mu g/L using three commercially available enzyme-linked immunosorbent assay (ELISA) kits. One kit, Human FGF-23[C-term] (Immutopics Inc, USA) measures total FGF-23 whereas the other two kits, Immutopics intact and FGF-23 ELISA (Kainos, Japan) are reported to measure only the biologically active intact molecule. Results We have detected a significant inverse correlation of -0.565 (P < 0.0001) between serum ferritin when < 50 mu g/L and FGF-23 using the C-terminal assay. This relationship is also shown with the Immutopics intact assay but is not demonstrated with the Kainos intact assay. Conclusion The measurement of FGF-23 by both Immutopics assays is altered in the presence of low circulating concentrations of serum ferritin whereas with the Kainos intact assay this effect was not demonstrated. Serum ferritin should be measured when an elevated FGF-23 is obtained using the Immutopics C-terminal or intact FGF-23 assay to prevent misdiagnosis of the cause of this abnormality.