4.7 Article

Sensitive determination of cadmium in water samples by room temperature ionic liquid-based preconcentration and electrothermal atomic absorption spectrometry

期刊

ANALYTICA CHIMICA ACTA
卷 628, 期 1, 页码 41-48

出版社

ELSEVIER
DOI: 10.1016/j.aca.2008.09.001

关键词

1-Butyl-3-methylimidazolium hexafluorophosphate; Room temperature ionic liquid; Cadmium; Preconcentration

资金

  1. Consejo Nacional de Investigaciones Cientificas y Tecnicas (CONICET)
  2. Agencia Nacional de Promocion Cientifica y Tecnologica (FONCYT) (PICT-BID)
  3. FONTAR [PMT II - CA1/073]

向作者/读者索取更多资源

A sensitive preconcentration methodology for Cd determination at trace levels in water samples was developed in this work. 1-Butyl-3-methylimidazolium hexafluorophosphate ([C4MIM][PF6]) room temperature ionic liquid (RTIL) was successfully used for Cd preconcentration, as cadmium -2-(5-bromo-2-pyridylazo)-5-diethylaminophenol complex [Cd-5-Br-PADAP]. Subsequently, Cd was back-extracted from the RTIL phase with 500 mu L of 0.5 mol L-1 nitric acid and deter-mined by electrothermal atomic absorption spectrometry (ETAAS). A preconcentration factor of 40 was achieved with 20 mL of sample. The limit of detection (LOD) obtained under optimum conditions was 3 ng L-1 and the relative standard deviation (R.S.D.) for 10 replicates at 1 mu g L-1 Cd2+ concentration level was 3.5%, calculated at peak heights. The calibration graph was linear from concentration levels near the detection limits up to at least 5 mu g L-1. A correlation coefficient of 0.9997 was achieved. Validation of the methodology was performed by standard addition method and analysis of certified reference material (CRM). The method was successfully applied to the determination of Cd in river and tap water samples. (C) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据