4.7 Article

A novel separation/preconcentration system based on solidification of floating organic drop microextraction for determination of lead by graphite furnace atomic absorption spectrometry

期刊

ANALYTICA CHIMICA ACTA
卷 623, 期 2, 页码 163-167

出版社

ELSEVIER
DOI: 10.1016/j.aca.2008.06.033

关键词

solidified floating organic drop microextraction; graphite furnace atomic absorption spectrometry; preconcentration; lead analysis; natural waters

向作者/读者索取更多资源

Solidified floating organic drop microextraction (SFODME) was successfully used as a sample preparation method for graphite furnace atomic absorption spectrometry (GFAAS). 20 mu L of 1-undecanol containing dithizone as the chelating agent (2 x 10(-4) mol L-1) was transferred to the water samples containing lead ions, and the solution was stirred for the prescribed time. The sample vial was cooled in an ice bath for 5 min. The solidified extract was transferred into a conical vial where it melted immediately, and then 10 mu L of it was analyzed by GFAAS. Factors that influence the extraction and complex formation, such as pH, concentration of dithizone, extraction time, sample volume, and ionic strength were optimized. Under the optimized conditions, a good relative standard deviation of +/- 5.4% at 10 ng L-1 and detection limit of 0.9ng L-1 were obtained. The procedure was applied to tap water, well water, river water and sea water, and accuracy was assessed through the analysis of certified reference water or recovery experiments. (C) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据