4.6 Article

Monitoring of pH changes in a live rat brain with MoS2/PAN functionalized microneedles

期刊

ANALYST
卷 143, 期 18, 页码 4469-4475

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c8an01149d

关键词

-

资金

  1. National Natural Science Foundation of China [21475034, 21505037, 21675041]

向作者/读者索取更多资源

Monitoring the dynamic pH changes in vivo remains very essential to comprehend the function of pH in various physiological processes. In this study, we report a high-performance electrochemical pH microneedle based on an acupuncture needle (AN) for real-time monitoring of pH changes in a rat brain. The pH microneedle was prepared by a layer-to-layer assembly of molybdenum disulfide (MoS2) nanosheets and polyaniline (PAN), with an attempt to achieve a highly sensitive detection of hydrogen ions (H+). The as-prepared PAN/MoS2/AN exhibited a high Nernstian response of -51.2 mV per pH over a wide pH range from 3.0 to 9.0, and excellent selectivity toward pH against other potential interfering species in the brain. Moreover, the corresponding open circuit potential rapidly increased and decreased when Na2CO3 or NaH2PO4 was injected into the rat brain, respectively, demonstrating that the PAN/MoS2/AN has an excellent response toward pH changes in vivo. This work provides a new potentiometric method for real-time monitoring of dynamic pH changes in vivo with high reliability and stability.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据