4.6 Article

Absolute quantification of protein copy number using a single-molecule-sensitive microarray

期刊

ANALYST
卷 139, 期 13, 页码 3235-3244

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c4an00091a

关键词

-

资金

  1. Single Cell Proteomics Project
  2. Proxomics Project of the Institute of Chemical Biology
  3. UK Engineering and Physical Sciences Research Council [EP/C54269X/1, EP/I017887/1]
  4. Engineering and Physical Sciences Research Council [1232251, EP/I017887/1] Funding Source: researchfish
  5. EPSRC [EP/I017887/1] Funding Source: UKRI

向作者/读者索取更多资源

We report the use of a microfluidic microarray incorporating single molecule detection for the absolute quantification of protein copy number in solution. In this paper we demonstrate protocols which enable calibration free detection for two protein detection assays. An EGFP protein assay has a limit of detection of <30 EGFP proteins in a microfluidic analysis chamber (limited by non-specific background binding), with a measured limit of linearity of approximately 6 x 10(6) molecules of analyte in the analysis chamber and a dynamic range of >5 orders of magnitude in protein concentration. An antibody sandwich assay was used to detect unlabelled human tumour suppressor protein p53 with a limit of detection of approximately 21 p53 proteins and a dynamic range of >3 orders of magnitude. We show that these protocols can be used to calibrate data retrospectively to determine the absolute protein copy number at the single cell level in two human cancer cell lines.