4.6 Article

Chemically modified flexible strips as electrochemical biosensors

期刊

ANALYST
卷 139, 期 18, 页码 4661-4672

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c4an00646a

关键词

-

资金

  1. CSIR, India

向作者/读者索取更多资源

A flexible and disposable strip sensor for non-enzymatic glucose detection is demonstrated in this work. The strips are prepared by using chemical modification processes followed by a simple electroless deposition of copper. Essentially, polyester overhead projector (OHP) transparent films are modified with a monolayer of 3-aminopropyltrimethoxysilane (APTMS) and polyaniline (PANI) conducting polymer. Later, nanostructured copper is deposited onto this modified film. Scanning electron microscope (SEM) and X-ray diffraction (XRD) studies are used for the structural, morphological and crystallinity characterization of the modified films. Electrochemical techniques, namely cyclic voltammetry (CV) and chronoamperometry (CA), are employed for the non-enzymatic detection of glucose. These studies clearly reveal the formation of homogeneous, close-packed spherical Cu particles converged into uniform film that exhibits a good catalytic activity towards the oxidation of glucose. The Cu/PANI/APTMS/OHP sensor displays a remarkable enhancement in the oxidation current density, a very high sensitivity value of 2.8456 mA cm(-2) per mM, and a linear concentration range from 100 mu M to 6.5 mM associated with glucose detection. Detection limit is estimated to be 5 mM and the response time of the sensor is determined to be less than 5 s. For comparison, similar studies are performed without PANI, namely Cu/APTMS/OHP films for glucose detection. In this case, a sensitivity value of 2.4457 mA cm(-2) per mM and a linear concentration range of 100 mu M-3 mM are estimated. The higher performance characteristics observed in the case of Cu/PANI/APTMS/OHP are attributed to the synergistic effects of the conducting polymer acting as an electron facilitator and the nanostructured Cu films. These disposable, flexible and low-cost strip sensors have also been applied to the detection of glucose in clinical blood serum samples and the results obtained agree very well with the actual glucose level.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据