期刊
ANALYST
卷 138, 期 9, 页码 2540-2545出版社
ROYAL SOC CHEMISTRY
DOI: 10.1039/c3an00133d
关键词
-
资金
- Science Foundation Ireland [08/SRC/B1412]
A technique for the in-process measurement of polymer stationary phase growth inside fused silica capillaries during the fabrication of monolithic porous layer open tubular (monoPLOT) columns is presented. In this work, capacitively coupled contactless conductivity detection (C4D) was applied as an online measurement tool for porous polymer layer growth within fused silica capillaries. The relationship between effective capillary diameter and C4D response was investigated for two polymers, butyl methacrylate-ethylene dimethacrylate (BuMA-EDMA) and polystyrene-divinylbenzene (PS-DVB) over a range of capillary diameters and layer thicknesses. The described technique can be used with both thermal and photo-initiated approaches for monoPLOT fabrication and provides an accurate, real-time measurement of the porous layer growth within the capillary, which should vastly improve column-to-column reproducibility. The technique was shown to be very precise, with a measured %RSD < 10%.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐