4.5 Article

Palladium biosensor

期刊

ELECTROANALYSIS
卷 19, 期 18, 页码 1909-1914

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.200703953

关键词

palladium ions; heavy metal; electrochemistry; electrochemical biosensor; environmental analysis

向作者/读者索取更多资源

In this paper we proposed a palladium(II) biosensor. The biosensor is based on determining of interactions between palladium(II) and metallothionein modified hanging mercury drop electrode by means of differential pulse voltammetry. We studied influence of two supporting electrolytes (potassium or sodium chloride) on the signals of the biosensor. Based on the results obtained we found potassium chloride (0.05 M) as the most suitable supporting electrolyte to determine palladium(II). The detection limit of the biosensor for palladium ions was evaluated as 100 nM with RSD about 10%. Moreover, we utilized the biosensor for measurement of the target molecule in the presence of human blood serum and human urine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据