A diamond-like carbon (DLC) coated quartz glass sample holder is used in a portable total reflection X-ray fluorescence (TXRF) spectrometer. A spot area of a water sample on the DLC sample holder becomes smaller than that on a quartz glass sample holder usually used in TXRF analysis because DLC is more hydrophobic. Therefore, the use of the DLC sample holder enhances fluorescent X-rays reaching a detector compared with the use of a quartz glass sample holder, leading to improvement in detection sensitivity. A detection limit of 28 pg is achieved for Cr in a river water sample when using the DLC sample holder.