4.5 Article

A search accelerated correct intensity Fourier transform microwave spectrometer with pulsed laser ablation source

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 78, 期 9, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.2786022

关键词

-

向作者/读者索取更多资源

The ablation of metal surfaces in the presence of a precursor gas produces reaction products which are often difficult to predict and highly dependent on ablation conditions. This article describes the successful development and implementation of a laser ablation source-equipped Fourier transform microwave spectrometer capable of observing 4 GHz regions of spectra in a single data acquisition event. The dramatically increased speed with which regions may be searched, when compared to other high resolution microwave techniques, allows the source conditions to be the prime variable in laser ablation microwave spectroscopic studies. A second feature of the technique is that observed spectral features have correct relative intensities. This is advantageous when assigning observed spectra. The study of two metal chlorides, AgCl and AuCl, illustrate the instrument's benefits. (c) 2007 American Institute of Physics.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。