4.6 Article

Polychromatic excitation improves detection limits in total reflection X-ray fluorescence analysis compared with monochromatic excitation

期刊

ANALYST
卷 135, 期 8, 页码 1909-1911

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c0an00009d

关键词

-

资金

  1. SENTAN, JST
  2. JSPS
  3. RIKEN

向作者/读者索取更多资源

Detection limits obtained by a portable total reflection X-ray fluorescence (TXRF) spectrometer with or without a monochromator are compared. A 1WX-ray tube (tube voltage: 20 kV) is used in this spectrometer. Polychromatic excitation improves the detection limits in TXRF analysis with the low power X-ray tube compared with monochromatic excitation. A detection limit of 26 pg is achieved for Co when using the weak polychromatic X-rays.