4.7 Article

Speciation and determination of thallium by on-line microcolumn separation/preconcentration by flow injection-flame atomic absorption spectrometry using immobilized oxine as sorbent

期刊

JOURNAL OF HAZARDOUS MATERIALS
卷 148, 期 1-2, 页码 446-452

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jhazmat.2007.02.059

关键词

T1 determination; T1 speciation; FI-FAAS; immobilized oxine

向作者/读者索取更多资源

A flow injection analysis (FIA) system incorporation a microcolumn of immobilized oxine on surfactant-coated alumina had been devised for performing rapid thallium enrichment/matrix removal in flame atomic absorption spectrometry (FAA). The preconcentration is based on the deposition of thallium(I) on microcolumn and subsequent elution with 500 mu l of sodium thiosulfate (I mol 1(-1)). In the presence of EDTA, only TI(I) was retained on the microcolumn. Total thallium was determined after reduction of Tl(III) to Tl(I) by hydroxyl amine hydrochloride. A sample volume of 25 ml resulted in a preconcentration factor of 77. Precision at 30 mu g l(-1)` was 2.6% RSD (n = 10). With 25 ml sampling volume a detection limit of 2.5 mu g l(-1) was determined. The effect of potential interfering ions on the determination was studied. The method was applied for the determination of thallium in water, waste water, hair, nail, coal, and standard reference alloys. The accuracy was assessed through recovery experiment, independent analysis by Furnace-AAS, and analysis of certified reference alloys. 0 2007 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据