ANALYTICA CHIMICA ACTA
卷 600, 期 1-2, 页码 6-13出版社
ELSEVIER
DOI: 10.1016/j.aca.2007.01.001
关键词
calibration; flow injection analysis; dilution method
A flow injection calibration system, originally designed and tested in our laboratory, is presented here as a versatile analytical tool serving for calibration purposes. It is characterized by a simple construction, easy operation and the possibility of offering rich measurement information with the use of a single standard solution. It is shown that depending on instrumental conditions and the composition of the calibration solutions the manifold is able to realize various calibration procedures according to interpolative, extrapolative and integration modes. As an experimental example, the determination of calcium in the cabbage sample by flame atomic absorption spectrometry has been calibrated by means of dilution method (DM) in the integration mode. (C) 2007 Elsevier B.V. All rights reserved.