4.3 Article

Sampling efficiency of a single-cell capillary electrophoresis system

期刊

CYTOMETRY PART A
卷 71A, 期 10, 页码 882-888

出版社

WILEY-LISS
DOI: 10.1002/cyto.a.20448

关键词

capillary electrophoresis; single-cell analysis; cell-sampling efficiency; cell lysis; pulsed lasers; green fluorescent protein

向作者/读者索取更多资源

Capillary electrophoresis (CE) combined with a laser-induced fluorescence (LIF) detection scheme is a powerful approach for single-cell analysis. For measurements requiring a high temporal resolution, CE-LIF is often combined with cell lysis systems based on pulsed lasers. Although extremely rapid, laser lysis has raised some concerns about the efficiency at which the cell contents are sampled. We have assembled a single-cell CE-LIF mounted on the stage of a microscope. This system was coupled with a nanosecond pulsed laser for cell lysis. We have analyzed green fluorescent protein (GFP) expressed in single mammalian cells and developed a novel approach to estimate the cell sampling efficiency (SE) based on the use of fluorescent calibration microspheres and flow cytometry. A significant advantage of this method is that it does not require any knowledge or assumption regarding the cell volume. We have evaluated the SE for different laser pulse energies (from 2 to 9 mu J) and two different pulse focal positions in the xy plane (0-10 mu m from the center of the cell). We found the maximum SE at the lowest energy (2 mu J), with the pulse focused directly on the cell. We have demonstrated the utility of a novel method to measure the SE of a single-cell CE system. The measurements presented in this study indicate that rapid cell lysis with nanosecond lasers requires careful optimization of pulse parameters for maximum sampling of the cell contents. (C) 2007 International Society for Analytical Cytology.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据