4.5 Article

Fabrication and mechanical testing of porous calcium phosphate bioceramic granules

期刊

出版社

SPRINGER
DOI: 10.1007/s10856-007-3128-0

关键词

-

向作者/读者索取更多资源

Porous hydroxyapatite/tricalcium phosphate (HA/TCP) granules were fabricated by a novel technique of vacuum impregnation of reticulated polyurethane (PU) foams with ceramic slip. The resultant granules had 5-10% interconnected porosity with controlled pore sizes necessary to allow bone ingrowth combined with good mechanical properties. Using PU foams with a different number of pores per inch (ppi), porous HA/TCP granules in the size range of 2-8 mm were successfully manufactured. Dieplunger tests were used to compare the compression and relaxation properties of the granules with those of a commercially available bone graft product, BoneSave(R). The results of the die-plunger testing showed that the experimental granules were stiffer than the BoneSave(R) materials and had less of a tendency to crumble to powder after testing. This therefore suggests that these experimental granules would be useful for impaction grafting and space filling applications.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据