4.5 Article

Accuracy of the Lifebox pulse oximeter during hypoxia in healthy volunteers

期刊

ANAESTHESIA
卷 68, 期 12, 页码 1220-1223

出版社

WILEY-BLACKWELL
DOI: 10.1111/anae.12382

关键词

-

资金

  1. NIGMS NIH HHS [K08 GM086511] Funding Source: Medline

向作者/读者索取更多资源

Pulse oximetry is a standard of care during anaesthesia in high-income countries. However, 70% of operating environments in low- and middle-income countries have no pulse oximeter. The Lifebox' oximetry project set out to bridge this gap with an inexpensive oximeter meeting CE (European Conformity) and ISO (International Organization for Standardization) standards. To date, there are no performance-specific accuracy data on this instrument. The aim of this study was to establish whether the Lifebox pulse oximeter provides clinically reliable haemoglobin oxygen saturation (SpO2) readings meeting USA Food and Drug Administration 510(k) standards. Using healthy volunteers, inspired oxygen fraction was adjusted to produce arterial haemoglobin oxygen saturation (SaO2) readings between 71% and 100% measured with a multi-wavelength oximeter. Lifebox accuracy was expressed using bias (SpO2-SaO2), precision (SD of the bias) and the root mean square error (Arms). Simultaneous readings of SaO2 and SpO2 in 57 subjects showed a mean (SD) bias of -0.41% (2.28%) and Arms 2.31%. The Lifebox pulse oximeter meets current USA Food and Drug Administration standards for accuracy, thus representing an inexpensive solution for patient monitoring without compromising standards.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据