Reproductivity and repeatability of central corneal thickness measurement in keratoconus using the rotating scheimpflug camera and ultrasound pachymetry

期刊

AMERICAN JOURNAL OF OPHTHALMOLOGY
卷 144, 期 5, 页码 712-718

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ajo.2007.07.021

关键词

-

向作者/读者索取更多资源

PURPOSE: To assess repeatability, reproducibility, and agreement of rotating Scheimpflug camera (Pentacam Oculus, Wetzlar, Germany) and ultrasound pachymetry in measuring central thickness of keratoconic corneas. DESIGN: Method,comparison study. METHODS: In 33 patients with keratoconus (one eye per patient), two examiners each used both pachymetric methods to measure central corneal thickness (CCT); in the same session, measurements then were repeated by examiner 1 (A.M.). The difference between two examiners, and between first and second measurements by examiner 1, with both methods and the difference beV,veenthe two pachymetric methods in measuring central thickness of keratoconic corneas were noted. RESULTS: With the rotating Scheimpflug camera, inter-examiner correlation was higher (intra,class correlation coefficient [ICC], 0.98 vs 0.76) and inter-examiner variability was lower (95% limits of agreement [95% LoA] -14.8 to 13.8 mu m vs -18.0 to +49.5 mu m) than with ultrasound pachymetry. Both methods showed close first-to second,measurement correlation (ICC, > 90), but the rotating Scheimpflug camera had lower variability (95% LoA, -14.5 to 14.2 mu m vs -27.4 to 26.0 mu m). Mean CCT was 478.9 +/- 34.6 mu m with the rotating Scheimpflug camera and 486.6 +/- 30 mu m with ultrasound pachymetry. Although the mean difference was small (-7.8 mu m), the 95% LoA (-43.8 to 28.2 mu m) showed that the difference between the two methods can be considerable. CONCLUSIONS: In keratoconic corneas, the rotating Scheimpflug camera provides measurements of central thickness that are more reproducible and repeatable than those obtained with ultrasound pachymetry. The rotating Scheimpflug camera seems to be suitable for disease staging and follow-up, when corneal thickness measurements may be repeated over time by different examiners.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据