4.7 Article

Measurement of visual evoked potential during and after periods of pulsed magnetic field exposure

期刊

JOURNAL OF MAGNETIC RESONANCE IMAGING
卷 26, 期 5, 页码 1353-1356

出版社

WILEY
DOI: 10.1002/jmri.21155

关键词

magnetic resonance imaging; safety; visual revoked potential; pulsed magnetic fields; P100

资金

  1. Medical Research Council [G9900259] Funding Source: Medline
  2. MRC [G9900259] Funding Source: UKRI
  3. Medical Research Council [G9900259] Funding Source: researchfish

向作者/读者索取更多资源

Purpose: To study the effect of switched magnetic fields used in MR scanners on the visual evoked potential (VEP) in human subjects. Materials and Methods: We have used an MRI gradient coil, remote, from an MRI magnet to produce a time-varying magnetic field (0.5 kHz, peak field similar to 8.7 T/second) in the human brain without the confounding effects of static field exposure, or accompanying acoustic noise. The VEP response to a 2-Hz reversal, 8 x 8 checkerboard, occupying 20 degrees of the visual field was recorded from occipital locations O1 and O2. VEP recordings were made every five minutes before, during, and after a 10-minute magnetic field exposure, period for seven subjects. Results: In contradiction to studies previously reported in the literature for fields of 50 Hz and 60 mT, no significant effects on the peak amplitude or latency of the VEP P100 O1 and O2 responses were found. Conclusion: Switched magnetic fields of a level and frequency comparable to those used in MRI do not have a significant effect on primary retinal or visual processing.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据