4.7 Article

Determination of benzene, toluene, ethylbenzene, xylenes in water at sub-ng l-1 levels by solid-phase microextraction coupled to cryo-trap gas chromatography-mass spectrometry

期刊

CHEMOSPHERE
卷 69, 期 9, 页码 1381-1387

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.chemosphere.2007.05.004

关键词

solid-phase microextraction; benzene; toluene; ethylbenzene; xylene; cryo-trap

向作者/读者索取更多资源

A trace analytical method of benzene, toluene, ethylbenzene and xylenes (BTEX) in water has been developed by using headspace solid-phase microextraction (HS-SPME) coupled to cryo-trap gas chromatography-mass spectrometry (GC-MS). The chromatographic peak shape for BTEX was improved by using cryo-trap equipment. The HS-SPME experimental procedures to extract BTEX from water were optimized with a 75 pm carboxen/polydimethylsiloxane (CAR/PDMS)-coated fiber at a sodium chloride concentration of 267 g 1(-1), extraction for 15 min at 25 degrees C and desorption at 290 degrees C for 2 min. Good linearity was verified in a range of 0.0001-50 mu g 1(-1) for each analyte (r(2) = 0.996-0.999). The limits of detection (LODs) of BTEX in water reached at sub-ng 1(-1) levels. LODs of benzene, toluene, ethylbenzene, m/p-xylene and o-xylene were 0.04, 0.02, 0.05, 0.01 and 0.02 ng 1-1, respectively. The proposed analytical method was successfully used for the quantification of trace BTEX in ground water. The results indicate that HS-SPME coupled to cryo-trap GC-MS is an effective tool for analysis of BTEX in water samples at the sub-ng 1(-1) level. (C) 2007 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据