4.4 Article

Signal suppression/enhancement in HPLC-ESI-MS/MS from concomitant medications

期刊

BIOMEDICAL CHROMATOGRAPHY
卷 21, 期 11, 页码 1143-1150

出版社

JOHN WILEY & SONS LTD
DOI: 10.1002/bmc.863

关键词

signal suppression; signal enhancement; HPLC-MS/MS; concomitant medication

向作者/读者索取更多资源

This paper presents a study of the signal suppression and enhancement effects in assays based on HPLC-ESI-MS/MS detection. The major focus was to investigate the effect of signal suppression/enhancement of typical co-administered (concomitant) medications, i.e. naproxen and ibuprofen. The results demonstrate that the analyte and internal standard can experience signal enhancement up to a factor of ca 2.9 if the test analyte or internal standard co-elute with concomitant. Experimental results also demonstrate that the analyte and internal standard signal increased by a factor of ca 2.0 in the negative ion mode at physiological relevant levels of naproxen (100 mu g/mL) and by a factor of ca 1.6 in the negative ion mode at physiological relevant level of ibuprofen (10 mu g/mL) in both neat and plasma samples. Signal enhancement significantly increased when concomitant medications ionized in the same ion mode as the analyte and internal standard. To overcome signal enhancement or potential suppression from concomitant medications, a comprehensive HPLC method needs to be developed with sufficient separation of concomitant medication from the analyte and internal standard. Other means to reduce signal enhancement or potential suppression include switching ionization polarity and performing comprehensive sample clean-up to remove concomitant medications before analysis. Copyright (c) 2007 John Wiley & Sons, Ltd.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据