4.2 Article

Combined impella and intra-aortic balloon pump support to improve both ventricular unloading and coronary blood flow for myocardial recovery: An experimental study

期刊

ARTIFICIAL ORGANS
卷 31, 期 11, 页码 839-842

出版社

BLACKWELL PUBLISHING
DOI: 10.1111/j.1525-1594.2007.00477.x

关键词

cardiac assist device; cardiogenic shock; cardiac unloading; intra-aortic balloon pump; hemodynamics; myocardial recovery

向作者/读者索取更多资源

Some patients in need of hemodynamic support do not respond to intra-aortic balloon pump (IABP) therapy. Hemodynamic stability can then be obtained by a more potent cardiac assist device, like the Impella catheter pump. Whether additional IABP support additional to Impella support can provide more optimal hemodynamic myocardial conditions is examined in this study. Seven sheep were implemented with IABP and Impella. An acute myocardial infarction was induced. Hemodynamic performance was assessed during baseline, during Impella support and IABP support individually, and during the combined Impella plus IABP support. The Impella support provided a reduction of afterload with 30% and an increase of coronary artery flow with 47%. The IABP increased coronary artery flow (13%), carotid artery flow (16%), and aortic ascending blood pressure (6%); a similar (but stronger) effect was provided when using the IABP support additional to Impella support and, respectively, increases of 33, 21, and 19% were established. The oxygen demand-supply ratio decreased by 25% due to the extra use of the IABP. A combination of IABP and Impella provides the most optimal hemodynamic myocardial conditions compared to either stand-alone support.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据