4.5 Article

Photoacoustic tomography of a rat cerebral cortex with a ring-based ultrasonic virtual point detector

期刊

JOURNAL OF BIOMEDICAL OPTICS
卷 12, 期 6, 页码 -

出版社

SPIE-SOC PHOTO-OPTICAL INSTRUMENTATION ENGINEERS
DOI: 10.1117/1.2823076

关键词

virtual point detector; ring detector; photoacoustic imaging; small animal imaging; aperture effect; resolution; signal-to-noise ratio

资金

  1. NIBIB NIH HHS [R01 EB000712] Funding Source: Medline
  2. NINDS NIH HHS [R01 NS046214, R01 NS46214] Funding Source: Medline

向作者/读者索取更多资源

We image a rat cerebral cortex in situ by using a ring-based ultrasonic virtual point detector developed previously. Compared to the image generated by a finite-aperture detector, the image generated by the virtual point detector has a uniformly distributed resolution throughout the imaged area, owing to the lack of aperture effect of the ultrasonic detector. At the periphery of the image, the signal-to-noise ratio of the image obtained by the virtual point detector is also better than that of a finite-aperture detector. Furthermore, the virtual point detector can be scanned inside the brain to improve the local signal-to-noise ratio. (C) 2007 Society of Photo-Optical Instrumentation Engineers.