3.8 Article

Validity of a newly developed ultraminiature cordless EMG measurement system

出版社

MOSBY-ELSEVIER
DOI: 10.1016/j.tripleo.2007.05.023

关键词

-

向作者/读者索取更多资源

To verify validity of a newly developed ultraminiature EMG measurement system ( BMS), the ability of BMS to record masseteric EMG was compared with that of a conventional polygraph system ( PG) in the daytime. Effective distance between the transmitter unit and receiver unit of BMS was also examined. Subjects were 12 healthy volunteers. During tapping, maximum clenching, and gum chewing of all subjects, distinct bursts were observed in EMG recorded by BMS as well as PG. RMS values of maximum clenching measured by BMS and PG showed a linear and significant correlation, and there was no significant difference between the data of BMS and PG. When distance between the transmitter unit and receiver unit of BMS was 100 cm or less, no artifact signal was observed. Having obtained these findings suggesting ability for precise measurement in the daytime, we are planning to use BMS in home sleep studies in the next step.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据