4.8 Article

Detection and quantification of anthrax lethal factor in serum by mass spectrometry

期刊

ANALYTICAL CHEMISTRY
卷 79, 期 22, 页码 8463-8470

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac701741s

关键词

-

向作者/读者索取更多资源

infection is a complex of protective antigen, which localizes the toxin to the cell receptor, and lethal factor (LF), a zinc-dependent endoproteinase whose known targets include five members of the mitogen-activated protein kinase kinase (MAPKK) family of response regulators. We have developed a method for detecting functional LF in serum. Anti-LF murine monoclonal antibodies immobilized on magnetic protein G beads were used to capture and concentrate the LF from serum. The captured LF was exposed to an optimized MAPKK based peptide substrate, which it hydrolyzed into two smaller peptides. The LF cleavage products were then analyzed by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MS) and quantified by isotope dilution-MS. The entire analytical method can be performed in less than 4 h with detection of LF levels as low as 0.05 ng/mL. The method was used to quantify LF levels in serum from rhesus macaques infected with B. anthracis. Serum samples obtained at day 2 postinfection contained 30-250 ng/mL LF and illustrated the clear potential to detect LF earlier in the infection cycle. This method represents a highly specific and rapid diagnostic tool for early anthrax and has a potential additional role as a research tool for understanding toxemia and effects of medical countermeasures for anthrax.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据