Use of manometric temperature measurement (MTM) and SMART™ freeze dryer technology for development of an optimized freeze-drying cycle

期刊

JOURNAL OF PHARMACEUTICAL SCIENCES
卷 96, 期 12, 页码 3402-3418

出版社

ELSEVIER SCIENCE INC
DOI: 10.1002/jps.20982

关键词

freeze drying/lyophilization; formulation; mathematical model; processing; proteins

向作者/读者索取更多资源

This report provides, for the first time, a summary of experiments using SMART (TM) Freeze Dryer technology during a 9 month testing period. A minimum ice sublimation area of about 300 cm 2 for the laboratory freeze dryer, with a chamber volume 107.5 L, was found consistent with data obtained during previous experiments with a smaller freeze dryer (52 L). Good reproducibility was found for cycle design with different type of excipients, formulations, and vials used. SMART (TM) primary drying end point estimates were accurate in the majority of the experiments, but showed an over prediction of primary cycle time when the product did not fully achieve steady state conditions before the first MTM measurement was performed. Product resistance data for 5% sucrose mixtures at varying fill depths were very reproducible. Product temperature determined by SMART (TM) was typically in good agreement with thermocouple data through about 50% of primary drying time, with significant deviations occurring near the end of primary drying, as expected, but showing a bias much earlier in primary drying for high solid content formulations (16.6% Pfizer product) and polyvinylpyrrolidone (40 kDa) likely due to water re-adsorption by the amorphous product during the MTM test. (c) 2007 Wiley-Liss, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据