4.5 Article

Parallel optical coherence tomography system

期刊

APPLIED OPTICS
卷 46, 期 34, 页码 8291-8297

出版社

OPTICAL SOC AMER
DOI: 10.1364/AO.46.008291

关键词

-

类别

资金

  1. NIBIB NIH HHS [EB01032] Funding Source: Medline

向作者/读者索取更多资源

We present the design and procedures for implementing a parallel optical coherence tomography (POCT) imaging system that can be adapted to an endoscopic format. The POCT system consists of a single mode fiber (SMF) array with multiple reduced diameter (15 mu m) SMFs in the sample arm with 15 mu m center spacing between fibers. The size of the array determines the size of the transverse imaging field. Electronic scanning eliminates the need for mechanically scanning in the lateral direction. Experimental image data obtained with this system show the capability for parallel axial scan acquisition with lateral resolution comparable to mechanically scanned optical coherence tomography systems. (c) 2007 Optical Society of America.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据