Direct volumetric blood flow measurement in coronary arteries by thermodilution

期刊

JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY
卷 50, 期 24, 页码 2294-2304

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jacc.2007.08.047

关键词

-

向作者/读者索取更多资源

Objectives This study sought to validate a new method for direct volumetric blood flow measurement in coronary arteries in animals and in conscious humans during cardiac catheterization. Background Direct volumetric measurement of blood flow in selective coronary arteries would be useful for studying the coronary circulation. Methods Based on the principle of thermodilution with continuous low-rate infusion of saline at room temperature, we designed an instrumental setup for direct flow measurement during cardiac catheterization. A 2.8-F infusion catheter and a standard 0.014-inch sensor-tipped pressure/temperature guidewire were used to calculate absolute flow (Q(thermo)) in a coronary artery from the infusion rate of saline, temperature of the saline at the tip of the infusion catheter, and distal blood temperature during infusion. The method was tested over a wide range of flow rates in 5 chronically instrumented dogs and in 35 patients referred for physiological assessment of a coronary stenosis or for percutaneous coronary intervention. Results Thermodilution-derived flow corresponded well with true flow (Q) in all dogs (Q(thermo) = 0.73 Q + 42 ml/min; R-2 = 0.72). Reproducibility was excellent (Q(thermo) = 0.96 X Q(thermo,2) + 3 ml/min; R-2 = 0.89). The measurements were independent of infusion rate and sensor position as predicted by theory. In the humans, a good agreement was found between increase of thermodilution-derived volumetric blood flow after percutaneous coronary intervention and increase of fractional flow reserve (R-2 = 0.84); reproducibility of the measurements was excellent (Q(thermo,1) = 1-0 Q(thermo,2) + 0.9 ml/min, R-2 = 0.97), and the measurements were independent of infusion rate and sensor position. Conclusions Using a suitable infusion catheter and a 0.014-inch sensor-tipped guidewire for measurement of coronary pressure and temperature, volumetric blood flow can be directly measured in selective coronary arteries during cardiac catheterization.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据