4.7 Article Proceedings Paper

Plasma processing for inducing bioactivity in stainless steel orthopaedic screws

期刊

SURFACE & COATINGS TECHNOLOGY
卷 202, 期 4-7, 页码 1242-1246

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.surfcoat.2007.07.075

关键词

plasma processing; orthopaedic screws; bioactive coatings; hydroxyapatite

向作者/读者索取更多资源

The work presented in this paper is centred on applying plasma processing for inducing bioactivity (ability of a material to bond with bone) in otherwise bioinert stainless steel screws commonly used in orthopaedic surgery. As-received cortical stainless steel screws were hydroxylated using a patented two-step plasma process developed by the authors. The bioactivity of the screws thus processed was investigated in vitro by treating them with simulated body fluid (SBF). The SBF treatment resulted in a thin, continuous surface layer of hydroxyapatite (HA, calcium hydroxy phosphate) on the screws, thus confirming their bioactivity. The plasma processed screws were also compared with wet chemically processed screws for the growth of HA from SBF, and it was found that the HA layer on the plasma processed screws exhibited higher (complete) substrate coverage. The realisation of bioactive screws, in particular fine-threaded stainless steel screws, offers a major advancement towards the fixation of both internal and external orthopaedic implants. (C) 2007 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据