4.6 Article

Two-axis magnetically-driven MEMS scanning catheter for endoscopic high-speed optical coherence tomography

期刊

OPTICS EXPRESS
卷 15, 期 26, 页码 18130-18140

出版社

OPTICAL SOC AMER
DOI: 10.1364/OE.15.018130

关键词

-

类别

资金

  1. PHS HHS [19768] Funding Source: Medline

向作者/读者索取更多资源

A two-axis scanning catheter was developed for 3D endoscopic imaging with spectral domain optical coherence tomography (SD-OCT). The catheter incorporates a micro-mirror scanner implemented with microelectromechanical systems (MEMS) technology: the micro-mirror is mounted on a two-axis gimbal comprised of folded flexure hinges and is actuated by magnetic field. The scanner can run either statically in both axes or at the resonant frequency (<= 350Hz) for the fast axis. The assembled catheter has an outer diameter of 2.8 mm and a rigid part of 12 mm in length. Its scanning range is +/- 20 degrees in optical angle in both axes with low voltages (1 similar to 3V), resulting in a scannable length of approximately 1 mm at the surface in both axes, even with the small catheter size. The catheter was incorporated with a multi-functional SD-OCT system for 3D endoscopic imaging. Both intensity and polarization-sensitive images could be acquired simultaneously at 18.5K axial scans/s. In vivo 3D images of human fingertips and oral cavity tissue are presented as a demonstration. (c) 2007 Optical Society of America.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据