4.6 Article Proceedings Paper

Comparing the tube and gel techniques for ABO antibody titration, as performed in three European centers

期刊

TRANSPLANTATION
卷 84, 期 12, 页码 S17-S19

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/01.tp.0000296019.85986.af

关键词

A/B antibody; A/B antibody titration; ABO antibody; ABO antibody titration; direct agglutination; immunoglobulin G (IgG); immunoglobulin M (IgM); indirect agglutination

向作者/读者索取更多资源

Data from 60 consecutive ABO-incompatible kidney transplantations performed in Stockholm, Sweden; Freiburg, Germany; and Uppsala, Sweden, revealed significant variation in preoperative A/B antibody levels, with median titers of 1:32, 1:128, and 1:8, respectively. We wanted to investigate whether these differences were method-related. The same samples from 21 healthy blood donors were analyzed in the three centers using current local methods. Results confirmed method-related differences, with higher A/B titers in Freiburg and lower titers in Uppsala compared with Stockholm. Results for the same sample differed by a median of three (range 0 to 6) titer steps. When the same number of samples were analyzed in the three centers using the same gel method and the same test erythrocytes, results differed by a median of one titer step (range 0 to 4) for the same sample. In conclusion, gel hemagglutination technique significantly decreases intercenter variation compared with tube technique.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据