Dose Reduction in Helical CT: Dynamically Adjustable z-Axis X-Ray Beam Collimation

期刊

AMERICAN JOURNAL OF ROENTGENOLOGY
卷 194, 期 1, 页码 W49-W55

出版社

AMER ROENTGEN RAY SOC
DOI: 10.2214/AJR.09.2878

关键词

CT; dose efficiency; helical CT; overscanning; radiation dose

资金

  1. National Institutes of Health [DK83007]
  2. NATIONAL INSTITUTE OF DIABETES AND DIGESTIVE AND KIDNEY DISEASES [P50DK083007] Funding Source: NIH RePORTER

向作者/读者索取更多资源

OBJECTIVE. The purpose of this study was to measure the dose reduction achieved with dynamically adjustable z-axis collimation. MATERIALS AND METHODS. A commercial CT system was used to acquire CT scans with and without dynamic z-axis collimation. Dose reduction was measured as a function of pitch, scan length, and position for total incident radiation in air at isocenter, accumulated dose to the center of the scan volume, and accumulated dose to a point at varying distances from a scan volume of fixed length. Image noise was measured at the beginning and center of the scan. RESULTS. The reduction in total incident radiation in air at isocenter varied between 27% and 3% (pitch, 0.5) and 46% and 8% (pitch, 1.5) for scan lengths of 20 and 500 mm, respectively. Reductions in accumulated dose to the center of the scan were 15% and 29% for pitches of 0.5 and 1.5 for 20-mm scans. For scan lengths greater than 300 mm, dose savings were less than 3% for all pitches. Dose reductions 80 mm or farther from a 100-mm scan range were 15% and 40% for pitches of 0.5 and 1.5. With dynamic z-axis collimation, noise at the extremes of a helical scan was unchanged relative to noise at the center. Estimated reductions in effective dose were 16% (0.4 mSv) for the head, 10% (0.8 and 1.4 mSv) for the chest and liver, 6% (0.8 mSv) for the abdomen and pelvis, and 4% (0.4 mSv) and 55% (1.0 mSv) for coronary CT angiography at pitches of 0.2 and 3.4. CONCLUSION. Use of dynamic z-axis collimation reduces dose in helical CT by minimizing overscanning. Percentage dose reductions are larger for shorter scan lengths and greater pitch values.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据