OBJECTIVE - We aimed to establish whether multi-nerve testing with a point-of-care nerve conduction device could be used to diagnose diabetic sensorimotor polyneuropathy. RESEARCH DESIGN AND METHODS - A total of 72 consecutive patients with diabetes underwent a full neurological examination and a concurrent evaluation for nine standard electrophysiological parameters using conventional nerve conduction studies (the reference standard) and a point-of-care device. RESULTS - Spearman coefficients for correlation of point-of-care and conventional parameters ranged between 0.76 and 0.91 (P < 0.001 in all comparisons). Agreement by the method of Bland and Altman was acceptable despite small systematic biases. Fifty subjects (69%) had neuropathy according to conventional criteria. The sensitivity and specificity for the point-of-care device to identify such neuropathy was 88 and 82%, respectively. CONCLUSIONS - A novel point-of-care device has reasonable diagnostic accuracy and thus may represent a sufficiently accurate alternative for detecting the diffuse electrophysiological criteria necessary to make the diagnosis of diabetic sensorimotor polyneuropathy.