4.5 Article

The utility and precision of analogue and digital preoperative planning for total hip arthroplasty

期刊

INTERNATIONAL ORTHOPAEDICS
卷 32, 期 3, 页码 289-294

出版社

SPRINGER
DOI: 10.1007/s00264-006-0317-2

关键词

-

向作者/读者索取更多资源

We prospectively compared the utility and precision of preoperative templating performed in printed films (analogue) with that performed on digital radiographs (digital) in 69 patients undergoing primary total hip replacement. Five patients were excluded when misplacement of the marker resulted in a magnification error greater that 10%; in the remaining patients (64 hips), the cup size was within +/- one size in 62 (97%) of the analogue and 52 (81%) of the digital (p=0.01) plans. The stem size was within +/- one size in 63 (98%) of the analogue and 60 (94%) of the digital (p=0.39) plans. The distance from the proximal corner of the lesser trochanter to the center of the prosthetic head (LTCD) in the analogue differed by 5 mm or more from the digital plan in nine cases (14%). Analogue preoperative planning yielded more predictable results than digital planning, particularly in terms of acetabular component size and LTCD that dictates limb lengthening-shortening. The sources of error were not clearly explained by variations in magnification. Inconsistent positioning of the magnification marker may jeopardise the safe implementation of digital templating.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据