An optical flow injection analysis (FIA) system for glucose measurement in tomato was constructed by using a glucose oxidase (GOD) immobilized reactor and a photodiode sensor, a peristaltic pump and a personal computer. The GOD reactor was crammed with strip-shaped of a GOD immobilized membrane to an acrylic pipe. The GOD was immobilized on polytetrafluoroethylene membrane by crosslinking with glutaraldehyde. Hydrogen peroxide generated by the GOD reaction oxidized luminol to produce chemiluminescence in the presence of horseradish peroxide. The chemiluminescence was detected by using the photodiode. The glucose measurement system was calibrated with standard glucose solutions. The calibration range of the system for glucose was 2.0-100.0 mmol L-1 stop. Measurement time for every sample was within 3 min. Glucose concentrations of a fresh and processed tomato samples measured with this glucose measurement system could be evaluated rapidly and non-laboriously.