4.3 Article

Interunit and Intraunit Reliability of the RT3 Triaxial Accelerometer

期刊

JOURNAL OF PHYSICAL ACTIVITY & HEALTH
卷 5, 期 4, 页码 527-538

出版社

HUMAN KINETICS PUBL INC
DOI: 10.1123/jpah.5.4.527

关键词

physical activity measurement; reproducibility; accelerometry

资金

  1. NCRR NIH HHS [M01 RR-0079] Funding Source: Medline
  2. NINR NIH HHS [R01NR008286] Funding Source: Medline

向作者/读者索取更多资源

Background: Interest in the quantification of physical activity is on the rise. Triaxial accelerometry has frequently been used; however, research on the reliability of these devices is limited. We examine the interunit and intraunit reliability of 22 RT3 triaxial accelerometers using a performance-documented laboratory agitator. Methods: The RT3 units were tested while moving in 2 directions (antero-posterior, medio-lateral) and speeds (150 and 275 RPM) on a shaker with simultaneous documented performance output for three 24-hour periods. Results: Minimal shaker variance was recorded for all trials (coefficients of variation [CVs] < 0.52%). Our data demonstrate good reliability within RT3s (CVs < 1.81%) but poor reliability among the 22 units (CVs range = 9.5% to 34.7%). Conclusions: In longitudinal studies, each subject should use the same RT3 unit at each assessment. The use of multiple RT3 units in cross-sectional studies is not recommended because data interpretation would be compromised by the high between-unit variability.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据