4.4 Article

Metal-on-metal hip implants:: do they impair renal function in the long-term?: A 10-year follow-up study

期刊

ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY
卷 128, 期 9, 页码 915-919

出版社

SPRINGER
DOI: 10.1007/s00402-007-0466-9

关键词

hip arthroplasty; metal-on-metal; serummetal-levels; cobalt; chrom

向作者/读者索取更多资源

Introduction The aim of our study was to investigate a potential influence of elevated serumcobalt and serumchromiumlevels on renal function at minimum 10 years after implantation of a metal-on-metal hip. Materials and methods Between November 1992 and June 1994 98 patients (44 m, 54 f) with an average age of 56 (22-79) years received a metal-on-metal bearing Metasul (TM). At the time of the 10-year follow-up, 15 patients had died and 8 were lost to follow-up. The remaining 75 patients had laboratory analysis including serumcreatinine and full blood cell count as well as chromium and cobalt serumlevels. Results Ten years postoperatively the median serumcreatinine level was 0.86 (0.55-1.51) mg/dl, the serumcreatinine clearance Ccr was in the normal range. The hemogram did not differ from that measured at the time of surgery. The median serumcobalt concentration was 0.75 (0.3-50.10) mu g/l and the serumchromium concentration was 0.95 (0.3-58.6) mu g/l, 10 years postoperatively. Conclusion Our long-term data do not show any influence of serum cobalt or chromium concentrations on renal function following total hip arthroplasty.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据