期刊
AMERICAN JOURNAL OF CARDIOLOGY
卷 103, 期 3A, 页码 20A-26A出版社
EXCERPTA MEDICA INC-ELSEVIER SCIENCE INC
DOI: 10.1016/j.amjcard.2008.11.019
关键词
-
This article discusses the advantages and disadvantages of methods for the measurement of platelet function. The focus is on tests that can be used to monitor antiplatelet activity in the setting of cardiovascular disease and potentially predict thrombosis and bleeding. The tests described are platelet aggregometry; impedance aggregometry; VerifyNow (Accumetrics, San Diego, CA); Plateletworks (Helena Laboratories, Beaumont, TX); platelet surface P-selectin, platelet surface-activated glycoprotein IIb/IIIa, and leukocyte-platelet aggregates; TEG Platelet Mapping system (Haemoscope, Niles, I L); Impact cone and plate(let) analyzer (DiaMed, Cressier, Switzerland); Platelet Function Analyzer-100 (Siemens Healthcare Diagnostics, Inc., Deerfield, IL); phosphorylation of vasodilator-stimulated phosphoprotein; serum thromboxane 132; and urinary 11-dehydro thromboxane B(2). Some of the factors that differentiate these tests are sample volume requirements, the use of whole blood, the presence of shear, point-of-care status, need for a technician, and expense. (C) 2009 Published by Elsevier Inc. (Am J Cardiol 2009;103[suppl]:20A-26A)
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐