期刊
AEROSOL SCIENCE AND TECHNOLOGY
卷 44, 期 8, 页码 706-712出版社
TAYLOR & FRANCIS INC
DOI: 10.1080/02786826.2010.482110
关键词
-
类别
资金
- EU [FP6-026140]
- Academy of Finland [1118615]
- U.S. Department of Energy Atomic Radiation Measurement (ARM)
- NOAA
During the Reno Aerosol Optics Study (RAOS) in June 2002, the prototype of the 3-wavelength Particle Soot Absorption Photometer (3 PSAP) was calibrated and a correction scheme was subsequently published (Virkkula, A., Ahlquist, N. C., Covert, D. S., Arnott, W. P., Sheridan, P. J., Quinn, P. K., and Coffman, D. J. (2005). Modification, Calibration and a Field Test of an Instrument for Measuring Light Absorption by Particles. Aerosol. Sci. Technol., 39:68-83). Unfortunately, the 3 PSAP flow rate had not been corrected during that analysis. The factors in the transmittance correction change after reanalysis with flow-corrected data. Reanalysis of the data also resolves an apparent discrepancy between the 1 PSAP and the 3 PSAP, and the two instruments now agree well.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐