Effect of Working Fluid on Sub-2 nm Particle Detection with a Laminar Flow Ultrafine Condensation Particle Counter

期刊

AEROSOL SCIENCE AND TECHNOLOGY
卷 43, 期 1, 页码 81-96

出版社

TAYLOR & FRANCIS INC
DOI: 10.1080/02786820802488194

关键词

-

资金

  1. NSF [ATM-0506674]

向作者/读者索取更多资源

The effect of working fluid on size-dependent activation efficiencies with the laminar flow ultrafine condensation particle counter described by Stolzenburg and McMurry (1991) was studied theoretically and experimentally. Criteria considered include tendency to avoid homogeneous nucleation within the condenser and toxicity. The working fluids that were identified have vapor pressures below that of butanol, so particles grow to smaller sizes and are more difficult to detect optically. Therefore we use a second, conventional CPC as a booster to grow particles to a detectable size. Experiments were performed to obtain the size-and material-dependent activation efficiencies for ethylene glycol, diethylene glycol, propylene glycol, oleic acid, and DOS. Using diethylene glycol and oleic acid, values of the 50% activation efficiency diameter, D-p50, for negatively charged particles generated by evaporating sodium chloride, ammonium sulfate, and silver were < 1.2 (< 0.8) nm, 1.4-1.5 (1.0-1.2) nm, and 1.9-2.0 (1.5) nm as mobility (mass) diameter, respectively. The stability of the UCPC for long-term operation using ethylene glycol and propylene glycol as working fluids was tested by monitoring the instrument's response to silver particles having size near Dp50 in an air stream at 40-45% relative humidity. The performance was steady (+/- 3%) for several days indicating the instrument performs stably during unattended operation for realistic atmospheric sampling conditions.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据