4.1 Article

Comparing the sperm concentration determined by the Makler and the Neubauer chambers

期刊

ACTAS UROLOGICAS ESPANOLAS
卷 32, 期 4, 页码 443-445

出版社

ENE EDICIONES SL
DOI: 10.1016/S0210-4806(08)73860-9

关键词

sperm concentration; Makler chamber; Neubauer chamber; fertile men; spermatozoa

向作者/读者索取更多资源

Objective: The aim of this work was to determine the concordance between the counts obtained with the Makler and the Neubauer chambers in ejaculates from fertile men. Design: Cross sectional study in Reproduction Group, University of Antioquia, Medellin-Colombia. Methods: 112 ejaculates from fertile men were analyzed in the present study. Sperm count was determined using both chambers, the comparisons between chambers were determined using a Bland-Altman plot. Results: The means values for sperm count were 107.8 and 106.2 x 10(6) sperm/mL using the Makler and the Neubauer chamber, respectively, and there was concordance between both methods. Conclusion: Our results show that the determination of sperm concentration with the Makler chamber is as accurate as with the Neubauer chamber and that either chamber can be used in routine semen analyses.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据