期刊
ACTA RADIOLOGICA
卷 50, 期 5, 页码 543-548出版社
SAGE PUBLICATIONS LTD
DOI: 10.1080/02841850902839700
关键词
Anatomy; CT; high resolution; digital radiography; ear; ear; nose; throat; head; neck
资金
- Paulo Foundation
- Otologic Research Foundation
- Biomedicum Helsinki Foundation
Background: Clinical cone-beam computed tomography (CBCT), used in diagnostics of dental and maxillofacial radiology for almost 10 years, allows three-dimensional (3D) imaging of a focused area, with reasonable radiation dose. Purpose: To clarify the applicability of CBCT in imaging of the temporal bone. Material and Methods: We imaged cadaver temporal bones, one non-operated and five postmortem operated, with CBCT to evaluate the accuracy of this method in showing clinically important landmarks and the positions of middle-ear implants. In addition, to clarify the imaging protocols for the best possible result, we conducted a contrast-to-noise ratio (CNR) analysis by imaging a specially built phantom insert with different protocols. Results: For all the temporal bones, image quality was good and of diagnostic value, and the surgical landmarks as well as positions and details of the implants could be accurately observed. Based on measurements conducted with the phantom, the best possible clarity of the images with the machine used (3D Accuitomo; Morita Co., Kyoto, Japan) was achieved with a tube voltage of 80 kVp and a current of 4 mA. Conclusion: Cone-beam CT is a promising new method for otologic imaging, based on its accuracy and relatively low radiation exposure per investigation.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据