4.5 Article

Reduced variability in cup positioning: the direct anterior surgical approach using navigation

期刊

ACTA ORTHOPAEDICA
卷 79, 期 6, 页码 789-793

出版社

TAYLOR & FRANCIS LTD
DOI: 10.1080/17453670810016867

关键词

-

向作者/读者索取更多资源

Background and purpose Correct positioning of the acetabular component is important in total hip arthroplasty (THA). We evaluated the effect of an imageless navigation system on the accuracy of cup positioning using a minimally invasive direct anterior approach. Methods Hip replacements were performed in 44 cadaveric hips (22 cadavers) that were divided randomly into 2 groups. In the study group, THA was performed using an imageless navigation system. In the control group, no navigation system was used. CT scans were taken postoperatively. Using 3D reconstructions, the cup position was determined in relation to the frontal pelvic plane. The goal was to place each cup at 45 of inclination and 20 of anteversion, as recommended by the manufacturer. Results In the navigated group, there was a statistically significantly smaller range of deviation from the target angles of inclination and anteversion relative to the control group. Interpretation We conclude that imageless navigation improves the accuracy of cup placement in minimally invasive THA using the direct anterior approach.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据