ACTA OPHTHALMOLOGICA
卷 91, 期 8, 页码 E595-E602出版社
WILEY
DOI: 10.1111/aos.12179
关键词
epiretinal implant; perceptual threshold; retinitis pigmentosa; stimulus-induced neuronal depression
资金
- EU [001837]
Purpose: As part of a clinical trial, an investigational epiretinal implant (IMI Intelligent Medical Implant) was implanted in a retinitis pigmentosa patient. The prosthesis was wirelessly controlled by a visual interface containing a microcamera, providing wireless data and energy transmission. Forty-nine (49) electrodes were used for pattern recognition. This study examined the changes of perceptual thresholds over time and its relation to long-term stimulation. The goal of the study was to introduce stimulus-related depression of perceptual threshold (StirDepth) measurements as a method to gain further insight into the safety profile of electrical stimulation. Methods: The perceptual threshold was defined as the level of stimulation intensity at which a phosphene perception with a probability of 50% was detected using the Best-PEST method. StirDepth was measured by comparing the threshold changes immediately before and after a stimulation session of using three active electrodes and one passive electrode, which served as control. Results: The initial threshold of the day remained stable over the observed period. In StirDepth measurement all thresholds raised significantly after the stimulation sessions. The threshold increase of the active electrodes never exceeded that of the inactive control electrode. Conclusions: StirDepth measurement is feasible in epiretinal implants. The prolonged stimulation raised no safety concerns in the patient. The threshold increase of both the active electrodes and the control electrode leads one to hypothesise that cognitive or neurophysiological effects are the cause rather than the desensitizing of the retinal network or incipient retinal damage.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据