期刊
APPLIED OPTICS
卷 50, 期 3, 页码 387-391出版社
OPTICAL SOC AMER
DOI: 10.1364/AO.50.000387
关键词
-
类别
资金
- National Basic Research Program of China [2011CB301802]
- National Natural Science Foundation of China (NSFC) [60736037, 61036005, 11004182, 10704070]
In this work we have replaced a commonly used Au film with an Au/Ag bimetallic film working as the sensing chip in the commercial surface plasmon resonance (SPR) instrument, Biacore 3000. The bimetallic chip gives out sharper SPR dip than that by the chip made of an Au film. The applicability and stability of the bimetallic chip on the Biacore 3000 are tested by real-time monitoring of biological coupling processes between sulfamethoxazole (SMX) and anti-SMX, which is useful in rapid detection of contaminations in food products. The bimetallic chip performs stably in 6 times' usage and regeneration. Our experiment is helpful to push the bimetallic film into real applications in commercial SPR instruments. (C) 2011 Optical Society of America
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐